13C One-Pulse Experiment
3/4/94
To optimize the standard One-Pulse experiment:
1. Collect a FID using the standard parameters.
2. Observe the FID and estimate the T2 relaxation time constant:
determine the time when 95% of the FID has decayed (essentially
the end of the FID), and divide by 3. See Measuring Relaxation
Time Constants.
3. Process the spectrum using the standard paramters.
4. If peaks are unintentionally folded, set sw=35000
and start again at step 1.
5. Using the interactive display, position the spectrum so that
the left-most and right-most peaks lie about 1 ppm from the spectrum
edges. Type movesw to automatically update sw and tof
.
6. Set np to 5056, 8256, 16448, 32832, or 65600. np should be ³ 2pT2sw. If you can't decide between two valuesfor np, choose the larger value. Use Table 1 as a guide:
A. Choose the column with sw higher than your sw determined from step 5
B. Choose the row with T2 larger than your T2 in the column selected in step A.
C. Set np to the value in the left column of the row selected
in step B.
Table 1. Relationship between sw, T2, and np. | |||||||
sw (hertz) | |||||||
np | 5000 | 7500 | 10000 | 15000 |
20000 | 25000 | 30000 |
2112 | £0.07 | £0.05 | £0.03 | £0.02 | £0.02 | £0.01 | £0.01 |
5056 | £0.16 | £0.11 | £0.07 | £0.05 | £0.03 | £0.02 | £0.02 |
8256 | £0.26 | £0.17 | £0.16 | £0.11 | £0.07 | £0.05 | £0.04 |
16448 | £0.52 | £0.27 | £0.26 | £0.17 | £0.13 | £0.10 | £0.07 |
32832 | £1.04 | £0.69 | £0.52 | £0.34 | £0.26 | £0.20 | £0.15 |
65600 | ³1.04 | ³0.69 | ³0.52 | ³0.34 | ³0.26 | ³0.20 | ³0.15 |
T2 (seconds) | |||||||
7. Set d1 » (T1_estimate) - (at).
T1_estimate can be estimated from table 2. d1 is an additional
delay for relaxation after the FID has been detected. For molecules
< 1000 a.m.u., d1 is approximated to be 0 if the previous steps
have been followed and you have selected np £
32832.
8. Type ernst(T1_estimate,90_pulse), where T1_estimate (in seconds) can be estimated from Table 2, and 90_pulse is the value of the 90° pulse listed at the spectrometer.
Table 2: T1 for optimized one-pulse experiment | |||
molecular weight | multiply T2 by | molecular weight | multiply T2 by |
£1000 a.m.u. | 1 | 2000 | 2.01 |
1200 | 1.05 |
4000 | 2.44 |
1400 | 1.12 |
6000 | 2.92 |
1600 | 1.42 |
8000 | 3.49 |
1800 | 1.71 |
³10,000 |
4.32 |
9. Set nt to a multiple of 4. For maximum signal-to-noise,
set nt to the largest number that time permits. Type time
to display the total experiment time. nt can be set to fill the
remainder of your spectrometer reservation by typing time(hours,minutes).
The spectrometer then calculates nt; round down this calculated
nt to the nearest multiple of 4. For instance, if I type time(0,15),
and the spectrometer sets nt to 362, I then type nt=360.
10. Set ss=2 pad=0
11. The standard 13C parameters decouple 1H.
To acquire 1H-coupled spectra to show 1H
multiplicities, type dm='yyn' dm='e' nt=nt*4.
12. Recheck all experimental parameters. Parameters listed below
are from standard parameters for 1H spectroscopy in
chloroform.
13. Type go or ga to start your experiment. Type
aa, sa, or halt to stop the experiment before
completion. Type ra to resume acquisition stopped with
sa.
Parameters shown by typing dg:
sfrq=100.559 spectrometer frequency
tn=C13 transmitter nucleus
at=1.312 acquisition time, change np and sw instead
np=65600 number of points, see step 6
sw=25000 spectral width, see step 5
fb=13800 filter band width, signals farther than fb from the center of the spectrum are
filtered out. Set by spectrometer based upon sw.
bs=16 block size, number of transients to collect before data is stored and processed
ss=0 or 2 steady state transients (no data collected) before nt acquisitions, see step 10
tpwr=55 transmitter power, ranges from 0 (no power) to 63 (maximum power)
pw=14.0 pulse width, see step 8
p1=0 not used in this experiment
d1=3.0 relaxation delay, see step 7
d2=0 not used in this experiment
tof=0 transmitter offset, adjusts sfrq to set center of spectrum, higher numbers moves
center to higher frequencies (higher ppm), see step 5
nt=100 number of transients, see step 9
ct completed transients, set by spectrometer at end of
acquisition
Decoupling Parameters
dn=H1 decoupler nucleus
dof=-426.0 decoupler offset frequency
dm=yyy decoupler mode, decoupler turned on for all three segments of this experiment
dmm=w decoupler modulation mode (w = waltz decoupling)
dmf=9900 decoupler modulation frequency
dpwr=30 decoupler power, ranges from 0 (no power) to 49
(maximum power)
Processing Parameters Relevant to Acquisition:
werr=wft performs function upon error, such as aa (abort experiment)
wexp=wft performs function (wft in this example) at end of experiment
wbs=wft performs function (wft in this example) at end of bs transients
wnt=wft performs function (wft in this example) at end
of nt transients
Flags:
il='n' interleaved experiment, not used in this experiment
in='n' checks sample spinning at beginning of experiment
dp='y' double precision
hs='nn' homospoil not used in tins experiment
temp=not used see Setting the Variable Temperature Unit
priority=5 used by spectrometer to prioritize tasks
Parameters not shown:
dw=0.0001136 dwell, time between collection of data points of FID, change sw instead.
pad=0 pre acquisition delay, not used in optimized experiment
rof2=10 alfa=20 sum of rof2+alfa is delay between pulse and acquisition
gain='n' gain automatically set by spectrometer
gain='y' allows manual setting of the gain
gain? displays gain setting
gain=x sets gain to value of x, ranging from
0 to 59.
Interesting Functions:
number of points: np ³ 2pT2sw
digital resolution: dr = sw/fn
dwell: dw = 1/2sw
acquisition time: at = dw*np = np/2sw ³ pT2
pulse width: pw = [90° pulse listed at spectrometer][cos-1{exp(-at/T1)}]
minimum line width at half-height = 1/pT2
A. Spectrum recorded with standard parameters.
B. Spectrum recorded with sw set too small. The TMS peak at 0 ppm has been folded and appears at 9.05 ppm with distorted phase.
C. Spectrum recorded with optimized sw and tof, as described in step 5.
D. Spectrum recorded with optimized sw and tof for the aliphatic
region. Normally the aromatic peaks would fold into the spectrum.
However, the filter bandwidth (fb) was automatically set to 900
Hz. Thus, any signals greater than 900 Hz from tof (greater than
4.50 ppm or less than -0.90 ppm) are strongly attenuated. Since
the benzene signals resonate 2364 Hz from tof, these folded peaks
are not visible. This spectrum is a very unique example.
A. sw=4400
tof=-426.0
B. sw=3800
tof=-426.0
C. sw=3300
tof=-964.3
fb=1900
D. sw=1473.9
tof=-1881.8
fb=900