DEPT Experiment
The DEPT Experiment creates Distortionless Enhancement by Polarization
Transfer. This experiment can be used to increase the signal
intensities of 13C, or can be used to distinguish quaternary,
methine, methylene, and methyl carbons.
1. Collect an optimized 13C spectrum.
2. Type gain='y' gain? --note the value of the gain
3. Type dept to set up the DEPT experiment.
4. Adjust the following parameters:
pw 90° pulse for 13C listed at console
tpwr transmitter power listed at console
pp 90° pulse for 1H ON DECOUPLER CHANNEL listed at console
pplvl pplvl listed at console
j average 1H-13C coupling constant (about 140 Hz)
d1 relaxation delay, approximately 3T1 - at
nt multiple of 4, best if multiple of 16
satdly to presaturate peak at center of spectrum, type satdly='y'
hs homospoil not recommended, type hs='nn'
5. To increase all 13C signal intensities, type mult=0.5
To distinguish protonation states, type mult=0.5,1.0,1.0,1.5
0.5 generates spectrum with only protonated carbons
1.0 generates spectrum with CH carbons, all other carbons suppressed
1.5 generates spectrum with CH and CH3 carbons pointed up and CH2 carbons down
0.5,1.0,1.0,1.5 generates spectra for autodept routine
6. Type ga to start the experiment.
7. If you used the DEPT experiment to increase all 13C signal intensities, see Data Processing 1D Spectra.
8. When the experiment is finished, type autodept to generate plots of unedited spectra, edited spectra, and automatic spectrum analysis. If some peaks are not automatically analyzed, manually perform the autodept routine:
A. Type wft and phase the spectrum
B. Set spectrum vertical scale (vs) and threshold (click on main menu, display, interactive, more, thresh) so that all peak heights are greater than the threshold.
C. Type fp adept ds(1) vsadj
D. Type vp=0 and then adjust vs
E. Type vs=vs/4 vo=vs ho=0 full dssa
F. Type ppa pll('all') pscale page to plot all spectra .
The results of the experiment are:
Spectrum #1: all carbons
Spectrum #2: CH carbons
Spectrum #3: CH2 carbons
Spectrum #4: CH3 carbons