IU NMR user guide

HMQC – Heteronuclear Multiple-Quantum Coherence Experiment

(Frank Gao, 8/1/01)

The HMQC experiment establishes correlations between protons and other nuclei such as 13C or 15N while detecting high sensitivity protons. The correlations between protons and heteronuclei proceed via double and zero quantum coherences.

Note: 1) On I500, you have to recable the spectrometer when you change the detected nucleus between proton and 13C (or other X nuclei). The other spectrometers do not require hardware changes. 2) The gradient version of HMQC/HMBC experiments is preferred if the spectrometer is equipped with a gradient accessory (e.g., I500). 3) The HMQC/HMBC experiments are best-done using the inverse probe.

1. Setting up the spectrometer

In order to minimize residual non-13C-bound proton signals and “t1-noise”, a number of operating conditions that can influence the stability of the system should be controlled: 1) run experiments non-spinning; 2) optimize the lock signal; 3) use proper RF filters; 4) use VT (variable temperature) regulation; 5) use a moderate air flow through the probe; and 6) run experiments at night or weekend if possible.

1.1 Setting air pressure, temperature, RF filter, etc.

  1. Set the air pressure to 20 PSI
  2. Use a 13C band-pass filter (e.g., BE109-22-8BB for I400 and VXR400) in the X channel (if you are not sure, ask the NMR staff for assistance)
  3. Set up a standard 1H experiment in expn (e.g., exp1)
  4. Type temp=25 spin=0 su                – set temperature and turn off the spinning
  5. Wait until the temperature displayed in the Acquisition Status Window is regulated at 25 ΊC

1.2 Locking and shimming

  1. Make sure the spinning is turned off.
  2. Lock and shim as usual (i.e., use lock power given on solvent sheet)
  3. Optimize the lock power:
  4. Set the lock gain such that the lock level is about 50.

2. Choosing the spectral windows

2.1 For 13C

  1. If on I500, make sure the spectrometer is cabled for 13C detection
  2. Join another experiment (e.g., exp2 by typing jexp2)
  3. Set up a standard 1D 13C experiment (note: use the same solvent name as for 1H)
  4. Enter temp=25
  5. Type nt=1 ga to acquire a quick 1D 13C spectrum. You should see the solvent peaks (unless the solvent contains no carbon, such as D2O).
  6. Type dscale to display the chemical shift scale. Reference the spectrum to the chemical shifts of solvent peaks (e.g., CDCl3 peaks should be at ~ 77 ppm).
  7. Narrow the 13C spectral window to leave about 10% on either side of the peaks of interest by using two cursors and typing movesw (although these peaks are usually not seen in the quick spectrum, they can be expected from a previously collected 13C spectrum for this sample or estimated from similar samples).
  8. Write down the tof and sw. They will be the dof and sw1, respectively, in the HMQC experiment.

2.2 For 1H

  1. If on I500, make sure the spectrometer is cabled for 1H detection
  2. Type jexp1                – go back to exp1
  3. Type nt=1 ga to collect a quick 1D 1H spectrum
  4. Narrow the 1H spectral window to leave about 10% on either side of the peaks of interest by using two cursors and typing movesw
  5. Move the parameters to the experiment in which HMQC will be performed (e.g., enter mp(1,3) to move the parameters from exp1 to exp3)

3. Calibrating the 1H pulse width (optional)

  1. Type jexp4                – join an experiment for the calibration
  2. Type mp(1,4)             – copy the 1H parameter set from exp1
  3. Enter temp=25
  4. Increase the relaxation delay d1 e.g., d1=10
  5. Enter pw=5 (or any other value less than 90Ί pulse width)
  6. Enter gain=’y’           – disable the auto-gain setting
  7. Type ga                     – acquire a spectrum
  8. Phase the spectrum
  9. Enter an array of pw values around the estimated 360Ί pulse width (i.e., 4*pw90, where pw90 is a pre-calibrated 90Ί pulse width for the standard sample). For example, enter pw=38.4, 39.2, 40, 40.8, 41.6, 42.4 assuming pw90=10.
  10. Type da                     – list the pw array you entered
  11. Type ga                     – acquire an array of spectra
  12. Type wft dssh            – display all spectra horizontally
  13. Find the value of pw for which the spectrum has minimum intensity, and divide this value by 4 to get the accurate 1H 90° pulse width for your sample.

4. Setting up a HMQC experiment

  1. Type jexp3                – join the experiment for HMQC
  2. Type iuhmqc             – read in the default parameter set for HMQC
  3. Enter temp=25
  4. Enter spin=0
  5. Make sure that the spinning is turned off
  6. Enter dof=…             tof from 13C spectrum that you wrote down in 2.1.8
  7. Enter sw1=…            sw from 13C spectrum that you wrote down in 2.1.8
  8. Make sure that the following parameters are set to their up-to-date values listed at the console:
  9. The following parameters may be changed upon your sample:

5. Optimizing the “null” delay (skip this if you deal with a macromolecule)

  1. Enter phase=1 nt=1 ni=1 ss=4 dm=’nnn’
  2. Enter an array of null values with one very short value (e.g., 0.001). For example, null=0.001, 0.1, 0.2, 0.3, 0.4, 0.5, 0.7, 1, 1.5, 2
  3. Type da                     – list the null array you entered
  4. Type ai                      – select the absolute-intensity display mode
  5. Type ga
  6. Type wft dssh
  7. Select the value of null for which either most of the peaks, or the biggest peaks, or the peaks you are most interested in, are approximately zero.

 

6. Presaturation set up (if desired)

Presaturation can be used to suppress the unwanted big peak (e.g., solvent peak, water peak) in a HMQC experiment. Set it up in the following manner:

  1. Enter null=0 ss=0
  2. Type ga                     – acquire a spectrum
  3. Type mf(4)                – copy the data in the current exp (exp3) to exp4
  4. Type jexp4                – join exp4
  5. Type wft
  6. Set the cursor to the peak that is to be removed by presatuation
  7. Type nl movetof
  8. Write down the value of tof. It will be the satfrq in the HMQC experiment
  9. Type jexp3                – go back to exp3 (for HMQC)
  10. Enter satfrq=…         the tof value you determined in the step 8
  11. Enter satmode=’y’
  12. Enter satdly=1.5       a significant time compared with T1 of the peak to be presaturated
  13. Enter d1=0.05
  14. Enter satpwr=10,7,4,1
  15. Type ga                     – acquire a series of spectra with different saturated power levels
  16. Type wft dssh
  17. Find the minimum satpwr for which the peak will be removed, and set satpwr to this value

7. Finalizing the set up and starting the HMQC experiment

  1. Reset the following parameters:
  2. Enter null=…           the optimized value determined in 5.7
  3. Type time                – estimate the experiment time
  4. You may change nt, and ni to suite your time, nt will cost you signal/noise, ni resolution in the 13C dimension
  5. Type au                   – start the HMQC experiment

8. Processing a HMQC spectrum

  1. Type gaussian         – set up the weighting function for the 1H dimension
  2. Type wft(1)              – Fourier transform the 1st FID for 1H dimensional phasing
  3. Phase this 1D spectrum in the usual way. Be aware that there may be some negative peaks in the spectrum. In some case, the 2nd FID is better suitable for phasing – just type wft(2) instead of wft(1).
  4. If the data acquisition is completed, type iuhmqcproc – automatically process the 2D HMQC data and display the 2D spectrum.
  5. If the data acquisition is in progress (but at least 4 FIDs have been acquired),
  6. Usually the phase correction on the 13C dimension (f1) is not required. In case that you need to phase the 13C dimension, follow the steps below:
  7. The same approach can also be used for f2 phasing with the 2D display in “F2 mode” (trace=’f2’)

9. Display and Plotting a 2D HMQC spectrum

9.1 Adjusting the vertical scale and threshold

9.2 Setting the references for a 2D spectrum

  1. Make sure you are in the interactive display mode – a pair of cursors (cross-line) or two pairs of cursors appear in the 2D display. To switch to the interactive display mode, click Redraw (if this button is in the menu) or click Main/Display/Color Map (or Contour).
  2. If two pairs of cursors are displayed (i.e., in the box mode), click the Cursor button in the interactive 2D display menu to change the display to the cursor mode (i.e., one pair of cursors are displayed).
  3. Move the F2 cursor to the position (usually a peak) to be referenced for the F2 dimension by clicking and dragging the mouse (left button), then type rl(x.xp) where x.x is the chemical shift in ppm (e.g., rl(7.27p)).
  4. Move the F1 cursor to the position to be referenced for the F1 dimension, then type rl1(x.xd) where x.x is the chemical shift in ppm (e.g., rl1(120.5d)).

9.3 Expanding a 2D display

9.4 Plotting a 2D spectrum

  1. Type full to adjust the position and size of a 2D spectrum to fill the entire screen (and page).
  2. Type dpcon(’pos’, 10, 1.5) to see how your 2D spectrum looks if plotted with 10 contours spaced 1.5 level apart. Try other numbers if you wish.
  3. The macro command iuplhxcor can be used for plotting a 2D HMQC spectrum (see the manual for this macro by typing man(’iuplhxcor’)), for instance
  4. Type pltext to plot the experimental text file.
  5. Type iupage (or click page) to send the plot to the printer.

 


Back to | IU NMR User Guide | NMR Facility Home Page |
Send comments to chemnmr@indiana.edu
Last updated: October 31, 2001
URL: http://nmr.chem.indiana.edu/NMRguide/2dexpt/HMQC.html
Copyright 1996, The Trustees of Indiana University