IU NMR user guide
HMQC Heteronuclear Multiple-Quantum Coherence Experiment
(Frank Gao, 8/1/01)
The HMQC experiment establishes correlations between
protons and other nuclei such as 13C or 15N while
detecting high sensitivity protons. The correlations between protons and
heteronuclei proceed via double and zero quantum coherences.
Note: 1) On I500, you have to recable the spectrometer when you change the detected
nucleus between proton and 13C (or other X nuclei). The other
spectrometers do not require hardware changes. 2) The gradient version of
HMQC/HMBC experiments is preferred if the spectrometer is equipped with a
gradient accessory (e.g., I500). 3) The HMQC/HMBC experiments are
best-done using the inverse probe.
1. Setting up the spectrometer
In order to minimize
residual non-13C-bound proton signals and t1-noise, a number
of operating conditions that can influence the stability of the system should
be controlled: 1) run experiments non-spinning; 2) optimize the lock signal;
3) use proper RF filters; 4) use VT (variable temperature) regulation; 5) use
a moderate air flow through the probe; and 6) run experiments at night or
weekend if possible.
1.1 Setting air pressure, temperature, RF filter, etc.
- Set the air pressure to 20 PSI
- Use a 13C band-pass filter (e.g., BE109-22-8BB for I400 and
VXR400) in the X channel (if you are not sure, ask the NMR staff for
assistance)
- Set up a standard 1H experiment in expn (e.g., exp1)
- Type temp=25 spin=0 su
set temperature and turn off the spinning
- Wait until the temperature displayed in the Acquisition Status Window is
regulated at 25 ΊC
1.2 Locking and shimming
- Make sure the spinning is turned off.
- Lock and shim as usual (i.e., use lock power given on solvent sheet)
- Optimize the lock power:
- For non-experienced user, increase the lock power by 3;
- For experienced user, find the highest lock power at which the lock is stable
and then lower the lock power by 3;
- Set the lock gain such that the lock level is about 50.
2. Choosing the spectral windows
2.1 For 13C
- If on I500, make sure the spectrometer is cabled for 13C
detection
- Join another experiment (e.g., exp2 by typing jexp2)
- Set up a standard 1D 13C experiment (note: use the same solvent
name as for 1H)
- Enter temp=25
- Type nt=1 ga to acquire a quick 1D 13C
spectrum. You should see the solvent peaks (unless the solvent contains no
carbon, such as D2O).
- Type dscale to display the chemical
shift scale. Reference the spectrum to the chemical shifts of solvent
peaks (e.g., CDCl3 peaks should be at ~ 77 ppm).
- Narrow the 13C spectral window to leave about 10% on either side of
the peaks of interest by using two cursors and typing movesw
(although these peaks are usually not seen in the quick
spectrum, they can be expected from a previously collected 13C
spectrum for this sample or estimated from similar samples).
- Write down the tof and sw.
They will be the dof and sw1, respectively, in the HMQC
experiment.
2.2 For 1H
- If on I500, make sure the spectrometer is cabled for 1H
detection
- Type jexp1
go back to exp1
- Type nt=1 ga to collect a quick 1D 1H spectrum
- Narrow the 1H spectral window to leave about 10% on either side of
the peaks of interest by using two cursors and typing movesw
- Move the parameters to the experiment in which HMQC will be performed (e.g.,
enter mp(1,3) to move the parameters from exp1 to exp3)
3. Calibrating the 1H pulse width (optional)
- Type jexp4
join an experiment for the calibration
- Type mp(1,4)
copy the 1H parameter set from exp1
- Enter temp=25
- Increase the relaxation delay d1 e.g., d1=10
- Enter pw=5 (or any other value less than 90Ί pulse width)
- Enter gain=y
disable the auto-gain setting
- Type ga
acquire a spectrum
- Phase the spectrum
- Enter an array of pw values around the estimated 360Ί pulse width
(i.e., 4*pw90, where pw90 is a pre-calibrated 90Ί pulse width
for the standard sample). For example, enter pw=38.4, 39.2, 40, 40.8, 41.6,
42.4 assuming pw90=10.
- Type da
list the pw array you entered
- Type ga
acquire an array of spectra
- Type wft dssh
display all spectra horizontally
- Find the value of pw for which the spectrum has minimum intensity, and
divide this value by 4 to get the accurate 1H 90° pulse width for
your sample.
4. Setting up a HMQC experiment
- Type jexp3
join the experiment for HMQC
- Type iuhmqc
read in the default parameter set for HMQC
- Enter temp=25
- Enter spin=0
- Make sure that the spinning is turned off
- Enter dof=
tof from 13C
spectrum that you wrote down in 2.1.8
- Enter sw1=
sw from 13C
spectrum that you wrote down in 2.1.8
- Make sure that the following parameters are set to their up-to-date values
listed at the console:
- pw
1H 90Ί pulse width
- tpwr
1H pulse power
- pwx
13C 90Ί pulse width
- pwxlvl
13C pulse power
- The following parameters may be changed upon your sample:
- j1xh
average 1JCH coupling constant (default 145 Hz)
- d1
relaxation delay (default 1.5 s)
5. Optimizing the null delay (skip this if you deal with a
macromolecule)
- Enter phase=1 nt=1 ni=1 ss=4 dm=nnn
- Enter an array of null values with one very short value (e.g.,
0.001). For example, null=0.001, 0.1, 0.2, 0.3, 0.4, 0.5, 0.7, 1, 1.5,
2
- Type da
list the null array you entered
- Type ai
select the absolute-intensity display mode
- Type ga
- Type wft dssh
- Select the value of null for which either most of the peaks, or the
biggest peaks, or the peaks you are most interested in, are approximately
zero.

6. Presaturation set up (if desired)
Presaturation can be used to
suppress the unwanted big peak (e.g., solvent peak, water peak) in a HMQC
experiment. Set it up in the following manner:
- Enter null=0 ss=0
- Type ga
acquire a spectrum
- Type mf(4)
copy the data in the current exp (exp3) to exp4
- Type jexp4
join exp4
- Type wft
- Set the cursor to the peak that is to be removed by presatuation
- Type nl movetof
- Write down the value of tof. It will be the satfrq in the HMQC
experiment
- Type jexp3
go back to exp3 (for HMQC)
- Enter satfrq=
the tof value you determined in the step 8
- Enter satmode=y
- Enter satdly=1.5 a significant time compared with T1 of the peak to be
presaturated
- Enter d1=0.05
- Enter satpwr=10,7,4,1
- Type ga
acquire a series of spectra with different saturated power levels
- Type wft dssh
- Find the minimum satpwr for which the peak will be removed, and set satpwr
to this value
7. Finalizing the set up and starting the HMQC experiment
- Reset the following parameters:
- phase=1,2
- dm=nny
- ss=32
number of steady-state scans
- nt=16
number of scans per FID, must be a multiple of 8
- ni=160
number of data points on the 13C dimension
- Enter null=
the optimized value determined in 5.7
- Type time
estimate the experiment time
- You may change nt, and ni to suite your time, nt will
cost you signal/noise, ni resolution in the 13C
dimension
- Type au
start the HMQC experiment
8. Processing a HMQC spectrum
- Type gaussian set up the weighting function for the 1H dimension
- Type wft(1)
Fourier transform the 1st FID for 1H
dimensional phasing
- Phase this 1D spectrum in the usual way. Be aware that there may be some
negative peaks in the spectrum. In some case, the 2nd FID is
better suitable for phasing just type wft(2) instead of wft(1).
- If the data acquisition is completed, type iuhmqcproc
automatically process the 2D HMQC data and display the 2D spectrum.
- If the data acquisition is in progress (but at least 4 FIDs have been
acquired),
- Enter gf1=n
disable the Gaussian function for the 13C dimension
- Type sinesq(2, xx, f1) set up a cosine-squared weighting function for
the 13C dimension. xx is the number of FIDs that have been acquired,
which can be found in the Acquisition Window, e.g.,
sinesq(2, 64, f1)
- Type wft2da
perform a 2D Fourier Transform
- Usually the phase correction on the 13C dimension (f1) is not required.
In case that you need to phase the 13C dimension, follow the
steps below:
- Make sure that the 2D spectrum is displayed in F1 mode display the
f1 axis (13C dimension in HMQC/HMBC) horizontally. If not, enter trace=f1
and click the Redraw button.
- Click the phase2D button, the trace to which the horizontal cursor is
positioned will be displayed on the top of the 2D spectrum.
- Click and drag the horizontal cursor up and down to select a trace which have
a peak toward the right-hand edge of the spectrum, then click the spec1 button
- Select one other trace which have a peak toward the left-hand edge of the
spectrum, then click the spec2 button
- Click the Phase button, the spectrum 1 (1st trace) will be
displayed on the screen. Click on the peak displayed near the right edge of
the spectrum and phase it. Do NOT click in the left part of the spectrum at
this time.
- Click the spectrum 2 button to display the 2nd trace. Click
the mouse near the right edge of the spectrum and do NOT rephase, then click
the mouse on the peak at the left edge of the spectrum and phase it.
- Click the spectrum 1 button to recheck the 1st trace. Repeat
the process again if necessary.
- Click the DONE and then Return button to quit from the 2D
phasing menu.
The same approach can also be used for f2 phasing with the 2D display in F2 mode
(trace=f2)
9. Display and Plotting a 2D HMQC spectrum
9.1 Adjusting the vertical scale and threshold
- Enter the value of the parameters vs2d (vertical scale) and th
(threshold) directly (e.g., vs2d=500 th=2), or
- Adjust them by properly using the mouse.
- A macro nm2d can be used to set up vs2d and th automatically.
9.2 Setting the references for a 2D spectrum
- Make sure you are in the interactive display mode a pair of cursors
(cross-line) or two pairs of cursors appear in the 2D display. To switch to
the interactive display mode, click Redraw (if this button is in the menu)
or click Main/Display/Color Map (or Contour).
- If two pairs of cursors are displayed (i.e., in the box mode), click the Cursor
button in the interactive 2D display menu to change the display to the cursor
mode (i.e., one pair of cursors are displayed).
- Move the F2 cursor to the position (usually a peak) to be referenced for the
F2 dimension by clicking and dragging the mouse (left button), then type
rl(x.xp) where x.x is the chemical shift in ppm
(e.g., rl(7.27p)).
- Move the F1 cursor to the position to be referenced for the F1 dimension, then
type rl1(x.xd) where x.x is the chemical shift in ppm
(e.g., rl1(120.5d)).
9.3 Expanding a 2D display
- In the interactive display mode, use the left mouse button to determine the
lower left corner of the region to be expanded and the right mouse button to
determine the upper right corner, then click expand
- To display full spectrum, click Full (in cursor mode)
9.4 Plotting a 2D spectrum
- Type full to adjust the position
and size of a 2D spectrum to fill the entire screen (and page).
- Type dpcon(pos, 10, 1.5) to
see how your 2D spectrum looks if plotted with 10 contours spaced 1.5 level
apart. Try other numbers if you wish.
- The macro command iuplhxcor can be used for plotting a 2D HMQC spectrum
(see the manual for this macro by typing man(iuplhxcor)), for instance
- iuplhxcor(pos, 10, 1.5, 0, 0) to plot the 2D spectrum solely with 10
contours spaced 1.5 level apart;
- iuplhxcor(pos, 10, 1.5, 1, 2) to plot the 2D spectrum with 1D
1H (in exp1) and 13C (in exp2) spectra
along the respective axes, those 1D spectra should be put in the respective
experiments previously;
- iuplhxcor(pos, 10, 1.5, 1, 0) to plot the 2D spectrum with 1D
1H spectrum (in exp1) only;
- iuplhxcor(pos, 10, 1.5, 0, 2) to plot the 2D spectrum with 1D
13C spectrum (in exp2) only.
- Type pltext to plot the experimental text file.
- Type iupage (or click page) to send the plot to the printer.

Back to | IU NMR User Guide |
NMR Facility Home Page |
Send comments to chemnmr@indiana.edu
Last updated: October 31, 2001
URL: http://nmr.chem.indiana.edu/NMRguide/2dexpt/HMQC.html
Copyright 1996,
The Trustees of Indiana University