Acquiring the experiment:
1. Join the experiment in which the 1H reference spectrum was collected. Type cosyps.
2. Type dg. Adjust the parameters listed below for the standard COSYPS:
sfrq=399.879 dm='n'
tn=H1 sw1 = sw from 1H reference spectrm
at = set by spectrometer ni = multiple of 64
np=2048 dp='n'
sw = sw from 1H reference spectrum hs='n'
ss=2
tpwr = listed at console
pw = 1H 90° pulse width
p1= 1H 90° pulse width
d1 = 1H relaxation delay (1-2 * 1H T1)
d2=0 rf1 = rf from 1H reference spectrum
presat='n' rfp1 = rfp from 1H reference spectrum
tof = tof from 1H reference spectrum wp1 = wp from 1H reference spectrum
nt = multiple of 4 sp1 = sp from 1H reference spectrum
ct = set by computer
Change ni, nt, and d1 to suit your scheduled time. The vertical (non-acquisition) resolution equals sw1/2ni. Typical values are listed below for sw1=4400:
d1 | ni | nt | time | vert. res | d1 | ni | nt | time | vert. res |
1.5 | 128 | 4 | 0:30 | 17.2 Hz |
1.5 | 128 | 64 | 8:01 | 17.2 Hz |
1.5 | 256 | 4 | 1:01 | 8.6 | 1.5 | 256 | 64 | 16:09 | 8.6 |
1.5 | 512 | 4 | 2:03 | 4.3 | 1.5 | 512 | 64 | 32:50 | 4.3 |
1.5 | 128 | 16 | 2:00 | 17.2 |
2.25 | 128 | 4 | 0:43 | 17.2 |
1.5 | 256 | 16 | 4:02 | 8.6 | 2.25 | 256 | 4 | 1:26 | 8.6 |
1.5 | 512 | 16 | 8:12 | 4.3 | 2.25 | 512 | 4 | 2:54 | 4.3 |
Processing, Displaying, and Plotting:
Process using the procedure for Processing a 2D Phase-Sensitive Experiment. Plot All Contours. The COSYPS spectrum may be symmeterized by typing foldt. Type acosy to perform automatic analysis of the COSYPS cross peaks.