How to Reboot the Spectrometers

Reboot the spectrometer radio frequency hardware console if:

To Reboot the Gem300:

Type gem300fix in the VNMR software window and follow the instructions. During the reboot procedure, teh spectrometer will ask you to push the console reset switch, which is located in the radio frequency hardware console---just open the console door.

To Reboot the I500 or I400:

  1. Open a UNIX window.
  2. Type su acqproc.
    The spectrometer will respond with killing acqproc
  3. Open the left door of the radio frequency hardware console. Locate the small, red RESET switch on th left-most computer board (this board is marked as the CPU board). Push in this RESET switch.
  4. Locate the Magnet/Sample REgulation board (7 slots to the right of the CPU board). The fail light should be RED. Wait until the fail light turns off and the READY AP light turns GREEN.
  5. Type su acqproc.
    The spectrometer will respond with starting acqproc.
  6. Type exit to close the UNIX window
  7. In the VNMR software window, type load='y' su
  8. Check the lock signal
  9. If the spectrometer still misbeahves, try this procedure again, then contact the NMR Facility Staff

To Reboot the VXR400:

  1. Open a UNIX window.
  2. Type su acqproc.
    The spectrometer will respond with killing acqproc
  3. Type su acqproc.
    The spectrometer will respond with starting acqproc.
  4. Type exit to close the UNIX window
  5. In the VNMR software window, type load='y' su
  6. Check the lock signal
  7. If the spectrometer still misbeahves, try this procedure again, then contact the NMR Facility Staff

Back to NMR Facility Users Guide
Back to NMR Facility Home Page