Lock Stability Test


Using a sample of 75 mg sucrose in 0.5 ml D2O and 250 mM NaCl, perform the following experiment:
90(H) - 1/2JCH - P1 - 90X(13C) - Acquire
P1 = 90(13C)
phase cycling: P1 (x, -x); receiver (x, -x)
Use a long recycle delay to allow for complete relaxation Two-scan experiments (with 4 dummy scans) should be performed, spearated by 20 second delays. A total of 60 such experiments should be performed on a weekday any time between 9am and 5pm.

Back to Map of NMR Facility
Back to NMR Facility Home Page
Send comments to chemnmr@indiana.edu
Last updated: July 30, 1997
URL: http://nmr.chem.indiana.edu/tour/specs/ls.html