PFG Profile Test
1H Test
Using gadolinium-doped D2O with H2O for a 1H line width of 2 Hz, perform
the following experiment:
90(H) - PFG - Acquire
The PFG is applied at 30 Gauss/cm for 2 msec. The test should be repeated
for each PFG amplifier connected to the Z-axis port of a PFG probe.
The amplitude of the water peak must be < 1% compared to the amplitude
observed without application of the PFG pulse.