Pulse Linearity Test
1H Test
Using gadolinium-doped D2O with H2O for a 1H line width of 2 Hz, perform
a standard 1D pulse-acquire experiment. Increment the pulse length
in 10 nsec increments from 0 to 1 microsecond. The signal must increase linearly over the range from 200 nsec to 1 microsecond. The r value is the
linearity of th signal vs pulse length; the effective start of pulse
is the extrapolation of this line to pulse length=0.