Pulse Phase Stability Test
1H Test
Using gadolinium-doped D2O with H2O for a 1H line width of 2 Hz, perform
the following experiment:
90(H) - D1 -90(H) - Acquire
D1 = 1.0 msec
transmitter frequency should be placed 37 Hz from the signal
Repeat 64 times
The standard deviation of the signal amplitude must be < 0.15%
Back to Map of NMR Facility
Back to NMR Facility Home Page
Send comments to chemnmr@indiana.edu
Last updated: December 29, 1997
URL: http://nmr.chem.indiana.edu/tour/specs/pps.html
Copyright 1997,
The Trustees of Indiana University