The NMR Facility staff have developed the following procedure to help you
independently generate high-quality NMR spectra. This procedure also helps you
to keep the spectrometer in good shape for you and the other 50+ users. THIS
IS NOT AN EXAM. Please consider this procedure as a method through which you
and the Facility staff can first interact to support your research with NMR
spectroscopy throughout your research at Indiana University
1. Complete the Request for NMR User Training
form and return it to the NMR Facility staff. You are STRONGLY encouraged NOT to
start training until you plan to routinely use the spectrometer for research.
You may NOT use the spectrometer before the first training session.
2. Schedule 1 or more training sessions with NMR Facility staff. At least one
training session with a NMR Facility staff member MUST FIRST be held, after
which you may schedule additional training sessions with the Facility staff
and/or you may also be trained by other experienced users.
3. Practice using the spectrometer with research samples between 8am and 5pm
weekdays (i.e., when a staff member is present in the Facility). If you plan
to practice using the spectrometer after 5pm or on weekends or holidays, you
MUST make prior arrangements for a staff member or a designated experienced
user to be present in the NMR Facility during your spectroscopy.
4. Collect, process, and plot a 1H spectrum of a typical research sample
within 15 minutes:
- Line shape much be symmetrical, spinning sidebands must be no more
than 2% of peak height. Write line width at 50% of the peak height and
(spinning sideband amplitude)/(peak height) on spectrum. Write ALL lock and
shim parameters, lock parameters, lock level, spin speed, and temperature on
back of the spectrum.
- Set the number of points so that digital resolution < 0.25 hz/point
after zero-filling.
- Set d1 to an appropriate value. Write on the back of the spectrum about
how you chose this value.
- Spectral window must have exactly 0.5 ppm between left edge of window and
left-most peak and exactly 1.0 ppm between right edge of window and right-most
peak.
- Set number of transients and steady state transients. Signal/noise of the
resulting spectrum must be a minimum of 10. Write the signal/noise on the
spectrum.
- Set line broadening to a good value to improve signal/noise without
significantly broadening your peaks.
- Spectrum must be referenced. Indicate which line was used for referencing
on spectrum.
- Plot the full spectrum, scale, title, normalized integration of three or
more peaks, and peak frequencies on one page.
- Plot ALL parameters with spectrum or on separate sheet.
- Save data.
5. Collect, process, and plot a 13C spectrum of a typical research sample
within 30 minutes:
- Line width at half-height must be 5.0 Hz, line shape must be
symmetrical, spinning sidebands must be no more than 2% of peak height. Write
line width at half-height and (spinning sideband amplitude)/(peak height) on
spectrum. Write ALL lock and shim parameters, lock level, spin speed, and
temperature on back of spectrum.
- Set the number of points so that digital resolution &al; 1 Hz/point after
zero-filling.
- Spectral window must have exactly 5 ppm between left edge of window and
left-most peak and exactly 10 ppm between right edge of window and right-most
peak.
- Set number of transients and steady state transients. Signal/noise of the
resulting spectrum must be a minimum of 5. Write the signal/noise on the
spectrum.
- Spectrum must be referenced. Indicate which line was used for referencing
on spectrum.
- Plot ALL parameters with spectrum or on separate sheet.
- Save data.
6. Make a list of at least 3 suggestions or questions that you have about the
spectrometer or NMR spectroscopy.
7. Review the spectra and your suggestion/question list with a staff member.
8. You're officially trained! Pick up your NMR spectroscopy merit badge,
collect more spectra, and publish!
In addition, we have offered and will continue to offer a NMR mini course in
July or August of each year. The lectures and demonstrations cover background
and practical aspects of taking NMR spectra in the NMR facility.
Back to NMR Facility Home page
Send comments to: chemnmr@indiana.edu
Last updated: Dec 29, 1997
URL: http://nmr.chem.indiana.edu/training.html
Copyright 1997,
The Trustees of Indiana University