| Pulse Amplitude Stability | VXR-400 | Gem-300 | I-400 | I-500 |
|---|---|---|---|---|
| 90 degree pulse | 1.30% | 0.10% | 0.12% | 0.09% |
| 30 degree pulse | 2.16% | 0.17% | 0.15% | 0.08% |
| 1 usec pulse | 2.15% | 0.21% | 0.27% | 0.15% |
| Pulse Phase Stability | VXR-400 | Gem-300 | I-400 | I-500 |
|---|---|---|---|---|
| standard deviation | 1.16% | 0.98% | 0.27% | 0.27% |
| Quadrature Image Test | VXR-400 | Gem-300 | I-400 | I-500 |
|---|---|---|---|---|
| 1 scan | 5.5% | 0.24% | 0.44% | 0.14% |
| 4 scans | 0.34% | 0.02% | 0.04% | 0.003% |
| 8 scans | 0.18% | 0.01% | 0.03% | 0.002% |
| 16 scans | 0.18% | 0.01% | 0.02% | 0.002% |
| Pulse Linearity | VXR-400 | Gem-300 | I-400 | I-500 |
|---|---|---|---|---|
| r value | 0.9992 | 0.9999 | 0.9999 | 0.9999 |
| effective start of pulse | 63.1 nsec | -150 nsec | -5.5 nsec | 2.32 nsec |
| PFG Profile | I-500 |
|---|---|
| X-axis PFG module | 0.011% |
| Y-axis PFG module | 0.018% |
| Z-axis PFG module | 0.0025% |
| Lock Stability | VXR-400 | Gem-300 | I-400 | I-500 |
|---|---|---|---|---|
| center band > satelite bands | X expts | X expts | X expts | 7 expts |
| center band > 2*satelite bands | X expts | X expts | X expts | 0 expts |
| Vibrations | VXR-400 | Gem-300 | I-400 | I-500 |
|---|---|---|---|---|
| amplitude | 0.475% | 0.38% | 0.052% | 0.029% |
| largest frequency | 6.0 Hz | 3.1 Hz | 4.9 Hz | 4.7 Hz |