I400 update

13C and 31P are ok, 1H and 19F work at reduced S/N and longer pulse widths. Probe default values have been updated but proton decoupling for 13C may be affected.

2D experiments should be run on the VXR400 and I500 instead of this instrument.

The I400 probe will be sent for repair soon!